HYGROFLEX5-EX – HF5-EX – TWO-CHANNEL TRANSMITTER
The new HygroFlex5-EX series is the latest development in two-channel transmitters for the exact measurement of humidity and temperature in explosive atmospheres. This device conforms to the latest international standards. The HF5-EX series consists of a robust aluminium transmitter with or without display. The attached probes are cast into a stainless steel tube, and are certified for operation in Zone 0/20. The transmitter itself is certified for Zone 1/21.
Characteristics:
• Measurement of relative humidity and temperature, optional output of dew points
and other calculated parameter
• Safe operation in areas with explosion risk
• Intrinsically safe probes
• Interchangeable stainless steel probes
• Robust aluminium housing
Matching probes: EX probes
Number of probe inputs | 1 |
---|---|
Accuracy at 23 ±5 °C | ±0.8 %rh / ±0.1 K |
Galvanic separation output signals | Yes |
Dimensions | 160 x 100 x 61 mm |
Operating range | -40…60 °C / -10…60 °C (Models with display) 0…100 %rh, non-condensing |
Display type | Optional graphics display. Presentation configurable. |
Backlight | without Backlight |
Trend indicator | Yes, for models with display |
Psychrometric calculations | All psychrometric values |
Analogue outputs scaleable by the user | Yes |
Software Update | With internet and HW4 software |
User information | from Software HW4 (devicemanager) |
Data processing via HW4 | Graphics, statistics, analysis, qualification, etc. |
Circuit type | 2 wire |
Service Interface | UART (internal in device) |
IP protection rating | IP66 |
FDA/GAMP compatibility | FDA 21 CFR Part 11 and GAMP compatible |
Measuring range with interchangeable probe | 0…100 %rh and -40…85 °C |
Measurement interval | 20…240 s |
Probe adjustment by software | Not via device menu (only outside the explosive zone with HW4 + AC3001) |
Analog output signal (standard) | 4…20 mA, two-wire |
Probe material | Stainless steel |
Enclosure material | Aluminium (DIN EN 1706 EN AC-AlSi 12 (Fe)) |
Cable connections | Connections: Ex-e terminals (0.2…2.5 mm2) Cable gland: 16 x 1.5 (Ø cable 4.5…7 mm) |
Weight | 1030 g / 1140 g |
Start-up time | Standard cold <60 s / warm <30 s |
Current consumption | 2×24 mA start up / 2×20 mA operation |
CE / EMC conform | 2014/30/EU |
Power supply | 10…28 VDC |
Accuracy analog output | <20 μA |
ATEX directives | 2014/34/EU (ATEX) CSA22.2, UL 508 |
EX identification | EX II 2(1) G Ex eb mb [ia Ga] IIC T5 Gb EX II 2(1) D Ex tb [ia Da] IIIC T80°C Db |
Identification EX North America | Class I, Div 1/2, Groups ABCD T5…T4, Class II, Div 1/2, Groups EFG T80°C…T110°C Ex e mb [ia Ga] IIC T5…T4 Ga/Gb Ex tb [ia Da] IIIC T80°C…T110°C Da/Db Class I, Zone 0/1, AEx e mb [ia Ga] IIC T5…T4 Ga/Gb Class II, Zone 20/21, AEx tb [ia Da] IIIC T80°C…T110°C Da/Db |